If you get the "Sorry, the file cannot be displayed. There seems to be a technical issue ..." error when trying to view an image, please make sure you access the site at http://w140.com/tekwiki/, not www.w140.com/tekwiki/ (with "www"). Sorry for the inconvenience - we are looking for a generic solution.

Difference between revisions of "410"

From TekWiki
Jump to: navigation, search
m
 
Line 35: Line 35:
 
Tek 410 5.jpg
 
Tek 410 5.jpg
 
Tek410.png | Catalog image
 
Tek410.png | Catalog image
Inerior front RSZ.jpg | Inerior front view
+
Inerior front RSZ.jpg | Interior front view
Inerior front wo CRT RSZ.jpg  | Inerior front view witout CRT assy
+
Inerior front wo CRT RSZ.jpg  | Interior front view witout CRT assy
Inerior top RSZ.jpg  | Inerior top view
+
Inerior top RSZ.jpg  | Interior top view
Inerior bottom RSZ.jpg | Inerior bottom view
+
Inerior bottom RSZ.jpg | Interior bottom view
Inerior back RSZ.jpg  | Inerior back view
+
Inerior back RSZ.jpg  | Interior back view
 
CRT assy RSZ.jpg | CRT assembly
 
CRT assy RSZ.jpg | CRT assembly
 
Vertical.jpg |‎ Vertical amplifier
 
Vertical.jpg |‎ Vertical amplifier

Latest revision as of 23:27, 9 October 2019

Tek 410 1.jpg

The Tektronix 410 Physiological Monitor is a specialized oscilloscope for medical applications.

It is a small, battery-powered instrument using a magnetically deflected CRT, providing ECG, EEG and pulse measurement modes.

The graticule includes a heart rate scale allowing readings in beats/min (at 50 mm/s).

The 067-0549-00 Input Adapter can be used to calibrate the 410.

Specifications

Sweep 25, 50 or 100 mm/s
Bandwidth 0.1 to 250 Hz (ECG, AUX) or 100 Hz (EEG)
Deflection 10 mm/50 μV (EEG), 20 mm/mV (ECG), 2 mm/mV (AUX)
Input impedance 2 MΩ (EEG, ECG) / 20 MΩ (AUX)
CMRR 150,000 : 1
Battery 10 size C NiCd cells, operating time 8-12 hours, charging 14-16 hours
CRT 5" magnetically deflected, P-7 phosphor